C1900 FAQ

 

      Question: Is there a way to reset the program to factory default?

      Answer: There is a factory reset command (Copy Standard) in the commissioning level. Once the <copy standard> command is activated, the unit will default to the original program when shipped from factory.

      Tip: It is generally a good idea to perform this step anytime an upgrade is installed. Also, if your instrument is experiencing problems (and programming appears correct), a <company standard> would be the last resort before sending unit in for service.

     

    Detailed Steps:

    1)

    Turn off the Security Link LK3, two pin jumper installed in lower position. (Processor Board: Top, Center)

    2)

    Press until you see

    3)

    Press first and hold, then press . You will see

    4)

    Press until you see

    5)

    Press once, you will see .

    6)

    Press once, if you have 2101 firmware, you will see .

    7)

    Press once, you will see , followed by .

    8)

    From here, you can simply exit the program by either recyling power, or by pressing   until you see , then press until you see . Now press  until you see  , then press to exit programming.

    9)

    Turn on the Security Link LK3, two pin jumper installed in upper position. (Processor Board: Top, Center)